[Publicación periódica]
Título : |
In vivo absolute quantification of carnosine in the vastus lateralis muscle with 1H MRS using a surface coil and water as internal reference |
Tipo de documento : |
documento electrónico |
Autores : |
Vega, Gloria, Autor ; Ricaurte, German, Autor ; Mauricio Estrada Castrillón, Autor ; Reyngoudt, Harmen, Autor ; Cardona, Oscar M., Autor ; Gallo Villegas, Jaime A., Autor ; Narváez Sánchez, Raúl, Autor ; Calderón, Juan C., Autor |
Fecha de publicación : |
2023 |
ISBN/ISSN/DL : |
98700011900 |
Idioma : |
Inglés (eng) |
Materias : |
Carnosine
|
Palabras clave : |
Carnosine Proton magnetic resonance spectroscopy Skeletal muscle Vastus lateralis Gastrocnemius |
Clasificación: |
AC-2023-001 |
Resumen : |
Phantom’s experiments showed a maximum error of 7%, highlighting the validity of the measurements in the study setup. The carnosine concentrations (mmol/kg ww, mean?±?SD) measured in the GM were 6.8?±?2.2 with the VC (CcarVC) and 10.2?±?3.0 with the SC (CcarSC) (P?=?0.013; n?=?9). Therefore, a correction was applied to these measurements (CcarVC?=?0.6582*CcarSC), to make coils performance comparable (6.8?±?2.2 for VC and 6.7?±?2.0 for SC, P?=?0.97). After that, only the SC was used to quantify carnosine in the VLM, where a concentration of 5.4?±?1.5 (n?=?30) was found, with significant differences between men (6.2?±?1.3; n?=?15) and women (4.6?±?1.2; n?=?15). The error in quantitation was 5.3–5.5% with both coils. |
Nota de contenido : |
To standardize a method for 1H MRS intramuscular absolute quantification of carnosine in the thigh, using a surface coil and water as internal reference. |
Derechos de uso : |
Permiso de republicación otorgado al HPTU |
En línea : |
https://link.springer.com/article/10.1007/s00256-022-04149-8 |
Enlace permanente : |
https://hospitalpablotobon.cloudbiteca.com/pmb/opac_css/index.php?lvl=notice_dis |
[Publicación periódica] In vivo absolute quantification of carnosine in the vastus lateralis muscle with 1H MRS using a surface coil and water as internal reference [documento electrónico] / Vega, Gloria, Autor ; Ricaurte, German, Autor ; Mauricio Estrada Castrillón, Autor ; Reyngoudt, Harmen, Autor ; Cardona, Oscar M., Autor ; Gallo Villegas, Jaime A., Autor ; Narváez Sánchez, Raúl, Autor ; Calderón, Juan C., Autor . - 2023. ISSN : 98700011900 Idioma : Inglés ( eng)
Materias : |
Carnosine
|
Palabras clave : |
Carnosine Proton magnetic resonance spectroscopy Skeletal muscle Vastus lateralis Gastrocnemius |
Clasificación: |
AC-2023-001 |
Resumen : |
Phantom’s experiments showed a maximum error of 7%, highlighting the validity of the measurements in the study setup. The carnosine concentrations (mmol/kg ww, mean?±?SD) measured in the GM were 6.8?±?2.2 with the VC (CcarVC) and 10.2?±?3.0 with the SC (CcarSC) (P?=?0.013; n?=?9). Therefore, a correction was applied to these measurements (CcarVC?=?0.6582*CcarSC), to make coils performance comparable (6.8?±?2.2 for VC and 6.7?±?2.0 for SC, P?=?0.97). After that, only the SC was used to quantify carnosine in the VLM, where a concentration of 5.4?±?1.5 (n?=?30) was found, with significant differences between men (6.2?±?1.3; n?=?15) and women (4.6?±?1.2; n?=?15). The error in quantitation was 5.3–5.5% with both coils. |
Nota de contenido : |
To standardize a method for 1H MRS intramuscular absolute quantification of carnosine in the thigh, using a surface coil and water as internal reference. |
Derechos de uso : |
Permiso de republicación otorgado al HPTU |
En línea : |
https://link.springer.com/article/10.1007/s00256-022-04149-8 |